GTT-2
In the GTT-2, blood flows under gravity during the whole measurement. The initial shear-rate is 12,000–14,000 s-1 . The rate of flow is expressed by the time-interval [d] between the consecutive blood drops. As shown in the graph below, initially [d] is small but increases as flow rate is reduced by the formation of thrombi in the gaps. When [d] reaches the level set in the instrument program (default 15 sec), “Occlusion Time OT” is displayed. The number of blood drops from OT until complete arrest of flow indicates thrombus stability at atmospheric pressure. There is a thrombus stabilization period (default 300 sec) during which the photosensors are inactive. After this stabilization period, the first drop of blood detected by the photosensor indicates the beginning of spontaneous thrombolysis (Lysis Time; LT). At atmospheric pressure the rate of thrombolysis is slow.